Line 30: Line 30:
  
   * Prior to using the sensor, use a check standard.   * Prior to using the sensor, use a check standard.
-  * If the reading of the check standard is outside of the accuracy acceptance limits, then recalibrate the sensor.+  * If the reading of the check standard is outside of the accuracy acceptance limits, then re-calibrate the sensor.
   * If the calibration is outside of the acceptance limits, then clean/maintain the sensor according to manufacturer recommendations. If maintenance doesn't help, then replace the sensor.   * If the calibration is outside of the acceptance limits, then clean/maintain the sensor according to manufacturer recommendations. If maintenance doesn't help, then replace the sensor.
  