To Consort website  |  Print page

User Tools

Site Tools


ph_calibration_requirement

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
ph_calibration_requirement [24/10/2018 07:02]
pieter
ph_calibration_requirement [24/10/2018 07:06]
pieter
Line 11: Line 11:
   * After measuring fluoride-containing solution and strong organic solution   * After measuring fluoride-containing solution and strong organic solution
   * There is a big difference between the temperature of the test sample and the temperature of the buffer solution that is used in the last calibration.   * There is a big difference between the temperature of the test sample and the temperature of the buffer solution that is used in the last calibration.
- 
-=== Simply check === 
-There is a simple way to quickly check if your pH meter needs a calibration at the moment. All you need to do is dip the pH probe into the calibration standard solutions such as pH 4.00 or pH 7.00, and see if the measurements have a big difference to the standard value. For most applications (if not for scientific research), if the difference is within 0.05 pH, for example the meter reads 6.97 in 7.00 solution or 4.03 in 4.00 solution, then it means the meter is in good condition and a calibration is not a must at the point. If the difference is over 0.05 pH, for example the meter reads 6.92 in 7.00 or 4.08 in 4.00, then a calibration is necessary and highly recommended before testing. 
  
 === Frequency === === Frequency ===
Line 25: Line 22:
   * How the sensor is maintained by the operator between each measurement.   * How the sensor is maintained by the operator between each measurement.
   * The customer experience in regard to samples and application.   * The customer experience in regard to samples and application.
 +
 +=== Simply check ===
 +There is a simple way to quickly check if your pH meter needs a calibration at the moment. All you need to do is dip the pH probe into the calibration standard solutions such as pH 4.00 or pH 7.00, and see if the measurements have a big difference to the standard value. For most applications (if not for scientific research), if the difference is within 0.05 pH, for example the meter reads 6.97 in 7.00 solution or 4.03 in 4.00 solution, then it means the meter is in good condition and a calibration is not a must at the point. If the difference is over 0.05 pH, for example the meter reads 6.92 in 7.00 or 4.08 in 4.00, then a calibration is necessary and highly recommended before testing.
  
 === General advice === === General advice ===
Line 30: Line 30:
  
   * Prior to using the sensor, use a check standard.   * Prior to using the sensor, use a check standard.
-  * If the reading of the check standard is outside of the accuracy acceptance limits, then recalibrate the sensor.+  * If the reading of the check standard is outside of the accuracy acceptance limits, then re-calibrate the sensor.
   * If the calibration is outside of the acceptance limits, then clean/maintain the sensor according to manufacturer recommendations. If maintenance doesn't help, then replace the sensor.   * If the calibration is outside of the acceptance limits, then clean/maintain the sensor according to manufacturer recommendations. If maintenance doesn't help, then replace the sensor.
  
ph_calibration_requirement.txt · Last modified: 24/10/2018 07:21 by pieter