====== ORP Electrode Maintenance ====== | ====== ORP Electrode Maintenance ====== | ||
+ | =====Before taking into use===== | ||
+ | When the electrode has a refillable reference chamber, make sure this chamber is always completely filled with the reference electrolyte solution (3…4 M KCl, article code B520 or B020). An electrode that has been stored dry must be immersed for at least ten hours in the reference electrolyte solution. | ||
+ | =====Storage===== | ||
+ | During short interruptions (e.g. storage) the electrode should be immersed in the electrolyte solution. In doing this it is always kept ready for use. When the interruption is longer than a month, use a storage bottle filled with the reference electrolyte solution and place it on the electrode tip in order to protect the glass bulb. | ||
=====ORP Electrode Cleaning===== | =====ORP Electrode Cleaning===== | ||
The following procedure will result in removal of many common contaminants from the platinum ORP electrode. Fouling of the electrode can, however, be deployment-specific, | The following procedure will result in removal of many common contaminants from the platinum ORP electrode. Fouling of the electrode can, however, be deployment-specific, |