To Consort website  |  Print page

User Tools

Site Tools


maintenance_ec

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
maintenance_ec [04/09/2018 10:16]
pieter [Methods]
maintenance_ec [12/10/2018 08:24]
pieter
Line 9: Line 9:
 ===== Methods ===== ===== Methods =====
 Use the Consort cleaning solution **B530** to clean the electrodes for general use. This action is preferably done on a regular time base to ensure a continuous good operation as much as possible. Use the Consort cleaning solution **B530** to clean the electrodes for general use. This action is preferably done on a regular time base to ensure a continuous good operation as much as possible.
-  * Place the with distilled water rinsed electrode in the cleaning solution for 10-15 minutes.+  * Place the with distilled water rinsed electrode in the cleaning solution for 10-15 minutes. Stir the electrode regularly.
   * Rinse it again with distilled water before storing the electrode.   * Rinse it again with distilled water before storing the electrode.
  
Line 17: Line 17:
   - For solutions containing organic fouling agents (fats, oils, etc.), clean probe with acetone (do not use it for synthetic material cells).   - For solutions containing organic fouling agents (fats, oils, etc.), clean probe with acetone (do not use it for synthetic material cells).
   - For algae and bacteria containing solutions, clean the probe with a bleach containing liquid.   - For algae and bacteria containing solutions, clean the probe with a bleach containing liquid.
-  - The ultimate solution to clean the measurement plates of the **Epoxy cell** in an effective way: dip a cotton swab in Toluene en gently wipe off the graphite measurement plates, rinse with water.+  - The ultimate solution to clean the measurement plates of the **Epoxy cell** in an effective way: dip a cotton swab in Toluene en gently wipe off the graphite measurement plates, rinse with distilled water.
 __Procedure 1 to 4__: Clean the cell by dipping or filling it with the cleaning solution and agitating for two or three minutes. __Procedure 1 to 4__: Clean the cell by dipping or filling it with the cleaning solution and agitating for two or three minutes.
  
maintenance_ec.txt · Last modified: 12/10/2018 08:24 by pieter