To Consort website  |  Print page

User Tools

Site Tools


maintenance_ec

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Last revision Both sides next revision
maintenance_ec [04/09/2018 10:16]
pieter [Methods]
maintenance_ec [04/09/2018 10:17]
pieter [Methods]
Line 17: Line 17:
   - For solutions containing organic fouling agents (fats, oils, etc.), clean probe with acetone (do not use it for synthetic material cells).   - For solutions containing organic fouling agents (fats, oils, etc.), clean probe with acetone (do not use it for synthetic material cells).
   - For algae and bacteria containing solutions, clean the probe with a bleach containing liquid.   - For algae and bacteria containing solutions, clean the probe with a bleach containing liquid.
-  - The ultimate solution to clean the measurement plates of the **Epoxy cell** in an effective way: dip a cotton swab in Toluene en gently wipe off the graphite measurement plates, rinse with water.+  - The ultimate solution to clean the measurement plates of the **Epoxy cell** in an effective way: dip a cotton swab in Toluene en gently wipe off the graphite measurement plates, rinse with distilled water.
 __Procedure 1 to 4__: Clean the cell by dipping or filling it with the cleaning solution and agitating for two or three minutes. __Procedure 1 to 4__: Clean the cell by dipping or filling it with the cleaning solution and agitating for two or three minutes.
  
maintenance_ec.txt · Last modified: 12/10/2018 08:24 by pieter