===== Methods ===== | ===== Methods ===== | ||
+ | Use the Consort cleaning solution **B530** to clean the electrodes for general use. This action is preferably done on a regular time base to ensure a continuous good operation as much as possible. | ||
+ | * Place the with distilled water rinsed electrode in the cleaning solution for 10-15 minutes. | ||
+ | * Rinse it again with distilled water before storing the electrode. | ||
+ | Also: | ||
- For most applications, | - For most applications, | ||
- For lime and other hydroxide containing solutions, clean with a 5-10% solution of hydrochloric acid (HCl). | - For lime and other hydroxide containing solutions, clean with a 5-10% solution of hydrochloric acid (HCl). |