maintenance_ec

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Next revision Both sides next revision
maintenance_ec [23/04/2018 07:20]
pieter [Methods]
maintenance_ec [23/04/2018 07:21]
pieter [Storage]
Line 13: Line 13:
   - For solutions containing organic fouling agents (fats, oils, etc.), clean probe with acetone (do not use it for synthetic material cells).   - For solutions containing organic fouling agents (fats, oils, etc.), clean probe with acetone (do not use it for synthetic material cells).
   - For algae and bacteria containing solutions, clean the probe with a bleach containing liquid.   - For algae and bacteria containing solutions, clean the probe with a bleach containing liquid.
-  - To clean the measurement plates of the **Epoxy cell** in an effective waydip a cotton swab in Toluene en gently wipe off the graphite measurement plates, rinse with water.+  - The ultimate solution to clean the measurement plates of the **Epoxy cell** in an effective waydip a cotton swab in Toluene en gently wipe off the graphite measurement plates, rinse with water.
 __Procedure 1 to 4__: Clean the cell by dipping or filling it with the cleaning solution and agitating for two or three minutes. __Procedure 1 to 4__: Clean the cell by dipping or filling it with the cleaning solution and agitating for two or three minutes.
  
Line 21: Line 21:
  
 ===== Storage ===== ===== Storage =====
-It is best to store cells so that the electrodes are immersed in deionised water, possibly enriched with a very small amount of household detergent. Rinse the electrode with distilled water before use.+It is best to store cells so that the electrodes are immersed in deionised water, possibly enriched with a small amount of household detergent. Rinse the electrode with distilled water before use.
  
 Any cell that has been stored dry should be soaked in distilled water for 5 to 10 minutes before use to assure complete wetting of the electrodes.  Any cell that has been stored dry should be soaked in distilled water for 5 to 10 minutes before use to assure complete wetting of the electrodes. 
maintenance_ec.txt · Last modified: 12/10/2018 08:24 by pieter