Line 5: | Line 5: | ||
* the calibration cannot be done successfully, | * the calibration cannot be done successfully, | ||
* the pH is correctly when measuring buffers but not when measuring some other reference solutions, | * the pH is correctly when measuring buffers but not when measuring some other reference solutions, | ||
+ | * the measurements are not repeatable, | ||
* ... | * ... | ||
- | * Check this list of possible reasons: [[Guide to troubleshooting|Guide to troubleshooting]] | + | Check this list of possible reasons: [[Guide to troubleshooting|Guide to troubleshooting]] |
+ | ===== Check electrode ===== | ||
Many measurement problems arise from the electrode or from a wrong use, rarely from the instrument itself. Therefore it is important to: | Many measurement problems arise from the electrode or from a wrong use, rarely from the instrument itself. Therefore it is important to: | ||
* Make sure there is no **air bubble trapped** in the pH sensitive bulb. If this should be the case, shake it like a clinical thermometer with the bulb downwards to remove it. | * Make sure there is no **air bubble trapped** in the pH sensitive bulb. If this should be the case, shake it like a clinical thermometer with the bulb downwards to remove it. | ||
Line 20: | Line 22: | ||
* When the electrode is fixed to the cable, it is difficult to test it. Look for very irregular behavior when moving the cable. | * When the electrode is fixed to the cable, it is difficult to test it. Look for very irregular behavior when moving the cable. | ||
* Check whether all contacts of the plugs are all bright and clean. | * Check whether all contacts of the plugs are all bright and clean. | ||
- | | + | |
+ | ===== Check instrument ===== | ||
+ | | ||
* Do not store the meter in cold humid environments. If so, and a measurement problem arises, please store it free of connections and packaging in a warm dry environment for a few days and try it again. | * Do not store the meter in cold humid environments. If so, and a measurement problem arises, please store it free of connections and packaging in a warm dry environment for a few days and try it again. | ||
* Follow the [[http:// | * Follow the [[http:// |