issues_with_ph_measurement

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Last revision Both sides next revision
issues_with_ph_measurement [23/09/2019 07:02]
pieter [pH Troubleshooting]
issues_with_ph_measurement [23/09/2019 07:10]
pieter [Check electrode]
Line 5: Line 5:
   * the calibration cannot be done successfully,    * the calibration cannot be done successfully, 
   * the pH is correctly when measuring buffers but not when measuring some other reference solutions,    * the pH is correctly when measuring buffers but not when measuring some other reference solutions, 
 +  * the measurements are not repeatable,
   * ...   * ...
-  * Check this list of possible reasons: [[Guide to troubleshooting|Guide to troubleshooting]]+Check this list of possible reasons: [[Guide to troubleshooting|Guide to troubleshooting]]
  
 +===== Check electrode =====
 Many measurement problems arise from the electrode or from a wrong use, rarely from the instrument itself. Therefore it is important to: Many measurement problems arise from the electrode or from a wrong use, rarely from the instrument itself. Therefore it is important to:
   * Make sure there is no **air bubble trapped** in the pH sensitive bulb. If this should be the case, shake it  like a clinical thermometer with the bulb downwards to remove it.   * Make sure there is no **air bubble trapped** in the pH sensitive bulb. If this should be the case, shake it  like a clinical thermometer with the bulb downwards to remove it.
Line 20: Line 22:
         * When the electrode is fixed to the cable, it is difficult to test it. Look for very irregular behavior when moving the cable.         * When the electrode is fixed to the cable, it is difficult to test it. Look for very irregular behavior when moving the cable.
         * Check whether all contacts of the plugs are all bright and clean.         * Check whether all contacts of the plugs are all bright and clean.
-  * Keep your instrument clean, **avoid moisture** on the connectors. + 
 +===== Check instrument ===== 
 +  * Keep your instrument clean, **avoid moisture** in any case, especially on the connectors. 
   * Do not store the meter in cold humid environments. If so, and a measurement problem arises, please store it free of connections and packaging in a warm dry environment for a few days and try it again.   * Do not store the meter in cold humid environments. If so, and a measurement problem arises, please store it free of connections and packaging in a warm dry environment for a few days and try it again.
   * Follow the [[http://www.consort.be/wiki/ph_measurement_tips#good_measurement_practices|Good Measurement Practices]] to avoid measurement errors.   * Follow the [[http://www.consort.be/wiki/ph_measurement_tips#good_measurement_practices|Good Measurement Practices]] to avoid measurement errors.
issues_with_ph_measurement.txt · Last modified: 23/09/2019 07:12 by pieter