Line 5: Line 5:
   * the calibration cannot be done successfully,    * the calibration cannot be done successfully, 
   * the pH is correctly when measuring buffers but not when measuring some other reference solutions,    * the pH is correctly when measuring buffers but not when measuring some other reference solutions, 
 +  * the measurements are not repeatable,
   * ...   * ...
 +Check this list of possible reasons: [[Guide to troubleshooting|Guide to troubleshooting]]
  
 +===== Check electrode =====
 Many measurement problems arise from the electrode or from a wrong use, rarely from the instrument itself. Therefore it is important to: Many measurement problems arise from the electrode or from a wrong use, rarely from the instrument itself. Therefore it is important to:
   * Make sure there is no **air bubble trapped** in the pH sensitive bulb. If this should be the case, shake it  like a clinical thermometer with the bulb downwards to remove it.   * Make sure there is no **air bubble trapped** in the pH sensitive bulb. If this should be the case, shake it  like a clinical thermometer with the bulb downwards to remove it.
Line 23: Line 26:
   * Follow the [[http://www.consort.be/wiki/ph_measurement_tips#good_measurement_practices|Good Measurement Practices]] to avoid measurement errors.   * Follow the [[http://www.consort.be/wiki/ph_measurement_tips#good_measurement_practices|Good Measurement Practices]] to avoid measurement errors.
  
-When an error arises, it is best to try to locate the cause first.  
  
 ====== Report a measurement problem ====== ====== Report a measurement problem ======