issues_with_ph_measurement [07/08/2019 07:07] pieter [pH Troubleshooting] |
issues_with_ph_measurement [23/09/2019 07:08] pieter |
||
---|---|---|---|
Line 5: | Line 5: | ||
* the calibration cannot be done successfully, | * the calibration cannot be done successfully, | ||
* the pH is correctly when measuring buffers but not when measuring some other reference solutions, | * the pH is correctly when measuring buffers but not when measuring some other reference solutions, | ||
+ | * the measurements are not repeatable, | ||
* ... | * ... | ||
+ | Check this list of possible reasons: [[Guide to troubleshooting|Guide to troubleshooting]] | ||
+ | ===== Check electrode ===== | ||
Many measurement problems arise from the electrode or from a wrong use, rarely from the instrument itself. Therefore it is important to: | Many measurement problems arise from the electrode or from a wrong use, rarely from the instrument itself. Therefore it is important to: | ||
* Make sure there is no **air bubble trapped** in the pH sensitive bulb. If this should be the case, shake it like a clinical thermometer with the bulb downwards to remove it. | * Make sure there is no **air bubble trapped** in the pH sensitive bulb. If this should be the case, shake it like a clinical thermometer with the bulb downwards to remove it. | ||
Line 23: | Line 26: | ||
* Follow the [[http:// | * Follow the [[http:// | ||
- | When an error arises, it is best to try to locate the cause first. | ||
====== Report a measurement problem ====== | ====== Report a measurement problem ====== |