Guide to troubleshooting for pH/ORP/Ion

When an error arises, it is best to try to locate the cause first. Here's a list of all possible reasons that might cause measurement issues:

READINGS ARE NOT REPRODUCIBLE

  • Is there sample carryover?
  • Are there sample interferences or complexing agents present?
  • Is the reference electrode junction contaminated?

SLOW RESPONSE (READINGS SLOWLY CHANGING)

  • Was the electrode stored in the wrong solution?
  • Was the electrode poisoned by the sample?

OUT OF RANGE READING

  • Is the electrode plugged into the controller properly?
  • Is there a reference electrode present?
  • Is there enough fill solution left in the reference electrode?
  • Is the electrode placed in the sample solution?

LOW SLOPE OR NO SLOPE

  • Are the standards too old, contaminated, or made wrong?
  • Has the sample pH been adjusted properly to the operating range of the electrode?
  • Is there an air bubble on electrode surface?
  • Is the controller okay?
  • Is there enough fill solution left in the reference electrode?

NOISY RESPONSE (READINGS RANDOMLY CHANGING)

  • Is the controller grounded?
  • Is there an air bubble on electrode surface?
  • Is the controller operating properly?
  • Is there enough fill solution left in the reference electrode?

DRIFTY RESPONSE (READINGS CONTINUOUSLY CHANGING)

  • Is there excessive leaking at the reference electrode junction?
  • Is the reference electrode junction clogged?
  • Was the sensing membrane poisoned by the sample?
  • Are there temperature problems?
  • Is the sample too concentrated?
  • Does the sensing membrane need conditioning?

INACCURATE READINGS (BUT CALIBRATION IS OK)

  • Are the standards incorrect?
  • Was the sample pH adjusted properly?
  • Is there sample carryover?
  • Are there sample interferences or complexing agents present?