To Consort website  |  Print page

User Tools

Site Tools


guide_to_troubleshooting

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
guide_to_troubleshooting [23/09/2019 07:05]
pieter [Guide to troubleshooting for pH/ORP/Ion]
guide_to_troubleshooting [18/10/2019 07:54]
pieter
Line 2: Line 2:
 When a measurement problem arises, it is best to try to locate the cause first. Here's a list of all possible reasons that might cause measurement issues: When a measurement problem arises, it is best to try to locate the cause first. Here's a list of all possible reasons that might cause measurement issues:
  
-===== READINGS ARE NOT REPRODUCIBLE  =====+===== Readings are not reproducible =====
   * Is there sample carryover?   * Is there sample carryover?
   * Are there sample interferences or complexing agents present?   * Are there sample interferences or complexing agents present?
   * Is the reference electrode junction contaminated?   * Is the reference electrode junction contaminated?
  
-===== SLOW RESPONSE (READINGS SLOWLY CHANGING)  =====+===== Slow response (Reading are changing slowly) =====
   * Was the electrode stored in the wrong solution?   * Was the electrode stored in the wrong solution?
   * Was the electrode poisoned by the sample?   * Was the electrode poisoned by the sample?
  
-===== OUT OF RANGE READING =====+===== Out of range =====
   * Is the electrode plugged into the controller properly?   * Is the electrode plugged into the controller properly?
   * Is there a reference electrode present?   * Is there a reference electrode present?
Line 17: Line 17:
   * Is the electrode placed in the sample solution?   * Is the electrode placed in the sample solution?
  
-===== LOW SLOPE OR NO SLOPE  =====+===== Low slope or no slope  =====
   * Are the standards too old, contaminated, or made wrong?   * Are the standards too old, contaminated, or made wrong?
   * Has the sample pH been adjusted properly to the operating range of the electrode?   * Has the sample pH been adjusted properly to the operating range of the electrode?
Line 24: Line 24:
   * Is there enough fill solution left in the reference electrode?   * Is there enough fill solution left in the reference electrode?
  
-===== NOISY RESPONSE (READINGS RANDOMLY CHANGING) =====+===== Noisy response (Readings are changing randomly) =====
   * Is the controller grounded?   * Is the controller grounded?
   * Is there an air bubble on electrode surface?   * Is there an air bubble on electrode surface?
Line 30: Line 30:
   * Is there enough fill solution left in the reference electrode?   * Is there enough fill solution left in the reference electrode?
  
-===== DRIFTY RESPONSE (READINGS CONTINUOUSLY CHANGING) ===== +===== Drifty response (Readings are changing continuously) =====
   * Is there excessive leaking at the reference electrode junction?   * Is there excessive leaking at the reference electrode junction?
   * Is the reference electrode junction clogged?   * Is the reference electrode junction clogged?
Line 38: Line 38:
   * Does the sensing membrane need conditioning?   * Does the sensing membrane need conditioning?
  
-===== INACCURATE READINGS (BUT CALIBRATION IS OK) ===== +===== Inaccurate readings (calibration is ok ...) =====
   * Are the standards incorrect?   * Are the standards incorrect?
   * Was the sample pH adjusted properly?   * Was the sample pH adjusted properly?
guide_to_troubleshooting.txt · Last modified: 18/10/2019 07:54 by pieter