Line 1: Line 1:
-====== Guide to troubleshooting ======+====== Guide to troubleshooting for pH/ORP/Ion ====== 
 +When a measurement problem arises, it is best to try to locate the cause first. Here's a list of all possible reasons that might cause measurement issues:
  
-When an error arises, it is best to try to locate the cause first. Here's a list of all possible reasons that might cause measurement issues: +===== READINGS ARE NOT REPRODUCIBLE  =====
- +
-READINGS ARE NOT REPRODUCIBLE +
   * Is there sample carryover?   * Is there sample carryover?
   * Are there sample interferences or complexing agents present?   * Are there sample interferences or complexing agents present?
   * Is the reference electrode junction contaminated?   * Is the reference electrode junction contaminated?
  
-SLOW RESPONSE (READINGS SLOWLY CHANGING) +===== SLOW RESPONSE (READINGS SLOWLY CHANGING)  =====
   * Was the electrode stored in the wrong solution?   * Was the electrode stored in the wrong solution?
   * Was the electrode poisoned by the sample?   * Was the electrode poisoned by the sample?
  
-OUT OF RANGE READING +===== OUT OF RANGE READING =====
   * Is the electrode plugged into the controller properly?   * Is the electrode plugged into the controller properly?
   * Is there a reference electrode present?   * Is there a reference electrode present?
Line 18: Line 17:
   * Is the electrode placed in the sample solution?   * Is the electrode placed in the sample solution?
  
-LOW SLOPE OR NO SLOPE +===== LOW SLOPE OR NO SLOPE  =====
   * Are the standards too old, contaminated, or made wrong?   * Are the standards too old, contaminated, or made wrong?
   * Has the sample pH been adjusted properly to the operating range of the electrode?   * Has the sample pH been adjusted properly to the operating range of the electrode?
Line 25: Line 24:
   * Is there enough fill solution left in the reference electrode?   * Is there enough fill solution left in the reference electrode?
  
-NOISY RESPONSE (READINGS RANDOMLY CHANGING) +===== NOISY RESPONSE (READINGS RANDOMLY CHANGING) =====
   * Is the controller grounded?   * Is the controller grounded?
   * Is there an air bubble on electrode surface?   * Is there an air bubble on electrode surface?
Line 31: Line 30:
   * Is there enough fill solution left in the reference electrode?   * Is there enough fill solution left in the reference electrode?
  
-DRIFTY RESPONSE (READINGS CONTINUOUSLY CHANGING) +===== DRIFTY RESPONSE (READINGS CONTINUOUSLY CHANGING) =====
   * Is there excessive leaking at the reference electrode junction?   * Is there excessive leaking at the reference electrode junction?
   * Is the reference electrode junction clogged?   * Is the reference electrode junction clogged?
Line 39: Line 38:
   * Does the sensing membrane need conditioning?   * Does the sensing membrane need conditioning?
  
-INACCURATE READINGS (BUT CALIBRATION IS OK) +===== INACCURATE READINGS (BUT CALIBRATION IS OK) =====
   * Are the standards incorrect?   * Are the standards incorrect?
   * Was the sample pH adjusted properly?   * Was the sample pH adjusted properly?