Line 23: Line 23:
   * Polluted electrode due to sticky products, deposits on the measurement plates, ... . Use an appropriate method to clean the products or deposits.   * Polluted electrode due to sticky products, deposits on the measurement plates, ... . Use an appropriate method to clean the products or deposits.
  
-==== These advices help also====+==== These advices help also ====
   * Water samples and products are subject to changes in conductivity due to evaporation. Close all containers immediately after measurement.    * Water samples and products are subject to changes in conductivity due to evaporation. Close all containers immediately after measurement. 
   * Use fresh and clean calibration solutions.    * Use fresh and clean calibration solutions. 
Line 75: Line 75:
   - Rinse the new cell, dip in __second__ standard and read the display: …………… (µS or mS ?)/cm   - Rinse the new cell, dip in __second__ standard and read the display: …………… (µS or mS ?)/cm
   - What is the temperature of the solutions? …………… °C   - What is the temperature of the solutions? …………… °C
-  - If any alarm occurs, what is EXACTLY shown on the display(s) ?+  - If any message occurs, what is EXACTLY shown on the display(s) ?
   * ………………………………………………………………………………………………………………   * ………………………………………………………………………………………………………………
   * ………………………………………………………………………………………………………………   * ………………………………………………………………………………………………………………