 ===== Measurement ranges ===== ===== Measurement ranges =====
-The choice of the cell constant also determines the measurement ranges of the instrument. Each range is multiplied with the value of the cell constant. This results e.g. for a device of the C3000, C6000 and R3600 ranges:+The choice of the cell constant also determines the measurement ranges of the instrument. Each range is multiplied with the value of the cell constant. This results e.g. for a device of:  
 + 
 +The C3000, C6000 and R3600 ranges:
 ^CC^Lowest^  ^  ^  ^ Highest| ^CC^Lowest^  ^  ^  ^ Highest|
 |1,0cm<sup>-1</sup>|0,00-20,00_µS/cm|0,0-200,0_µS/cm|0-2000_µS/cm|0,00-20,00_mS/cm|0,0-200,0_mS/cm| |1,0cm<sup>-1</sup>|0,00-20,00_µS/cm|0,0-200,0_µS/cm|0-2000_µS/cm|0,00-20,00_mS/cm|0,0-200,0_mS/cm|