1. The condition of the electrode' | 1. The condition of the electrode' | ||
* The correct electrolyte solution should be used! Use only the solution that is supplied with the electrode or the membrane kit (article code SZ02K). | * The correct electrolyte solution should be used! Use only the solution that is supplied with the electrode or the membrane kit (article code SZ02K). | ||
- | * The membrane (Teflon) should be intact, use a magnifying glass to check for visual damage, replace when required. | + | * The membrane (Teflon) should be intact, use a magnifying glass to check for visual damage, replace when required. If not sure, try with a new membrane. |
* The plastic body and the screw threads should be clean and intact. | * The plastic body and the screw threads should be clean and intact. | ||
* The O-ring should be present, intact and correctly placed. | * The O-ring should be present, intact and correctly placed. |