do_probe [04/04/2018 07:13] pieter [SZ10T Consort Dissolved Oxygen electrode] |
do_probe [04/04/2018 07:14] pieter |
||
---|---|---|---|
Line 53: | Line 53: | ||
Saturate distilled water in a vessel by purging air in it for approximately 20 minutes to create 100% air saturation. Perform the saturation of the solution under operating pressure. Allow the output signal to stabilise. Follow the instructions of the meter to calibrate the electrode at 100% saturation. | Saturate distilled water in a vessel by purging air in it for approximately 20 minutes to create 100% air saturation. Perform the saturation of the solution under operating pressure. Allow the output signal to stabilise. Follow the instructions of the meter to calibrate the electrode at 100% saturation. | ||
+ | ===== Troubleshooting ===== | ||
+ | |||
+ | http:// | ||