Line 20: Line 20:
   * Unscrew the membrane cap (8) from the electrode body (2).    * Unscrew the membrane cap (8) from the electrode body (2). 
   * Clean it with distilled water (DI water).    * Clean it with distilled water (DI water). 
-  * Fill the membrane cap with the electrolyte to a level above the membrane cartridge using the syringe provided. Hold the electrode in an upright position and gently screw the cap back onto the body. Make sure there remain no air bubbles in it.+  * Fill the membrane cap with the electrolyte (R001069, SZ02K, NaOH/Glycol solution) to a level above the membrane cartridge using the syringe provided. Hold the electrode in an upright position and gently screw the cap back onto the body. Make sure there remain no air bubbles in it.
   * Please inspect the membrane for tears or leakage. The membrane should be uniformly stretched across the inner body. Replace the membrane if any damage has occurred.   * Please inspect the membrane for tears or leakage. The membrane should be uniformly stretched across the inner body. Replace the membrane if any damage has occurred.
   * Let the electrode rest for a while to polarise. This can be followed by reading the output value with the meter. It is required to wait until the measurement has stabilised. This can take several hours.   * Let the electrode rest for a while to polarise. This can be followed by reading the output value with the meter. It is required to wait until the measurement has stabilised. This can take several hours.