To Consort website  |  Print page

User Tools

Site Tools


d230info

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Next revision
Previous revision
d230info [30/01/2015 09:12]
pieter created
d230info [23/02/2015 15:28] (current)
pieter [Requirements]
Line 1: Line 1:
 ====== D230 ====== ====== D230 ======
  
-Specifications: [[http://www.consort.be/Downloads/Documentation/D230_pg.pdf|Consort D230]]+===== Specifications =====
  
-Software: [[d230|D230]]+[[http://www.consort.be/Downloads/Documentation/D230_pg.pdf|Consort D230]] 
 + 
 +===== Requirements ===== 
 + 
 +- The system needs to be connected to a computer with a true **RS232** serial port. It does not function correctly with converters to USB or to Ethernet. When required to obtain a serial port card, check that it is based upon UART chips such as 16550, 16650, 16750 or 16950. 
 + 
 +- Use the RS232 **cable** which was supplied with the system. When required to obtain a new one, check that it is a 'straight forwarded' type. Do not use null-modem cables. 
 + 
 +- The software runs under **Windows** but has not been completely tested for Windows 8. Under Windows 7 needs to the compatibility to be set to 'Win XP SP3'
 + 
 +===== Software ===== 
 + 
 +[[d230|D230]]
d230info.1422609122.txt.gz · Last modified: 30/01/2015 09:12 by pieter