To Consort website  |  Print page

User Tools

Site Tools


d230info

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
d230info [30/01/2015 09:13]
pieter
d230info [23/02/2015 15:28] (current)
pieter [Requirements]
Line 9: Line 9:
 - The system needs to be connected to a computer with a true **RS232** serial port. It does not function correctly with converters to USB or to Ethernet. When required to obtain a serial port card, check that it is based upon UART chips such as 16550, 16650, 16750 or 16950. - The system needs to be connected to a computer with a true **RS232** serial port. It does not function correctly with converters to USB or to Ethernet. When required to obtain a serial port card, check that it is based upon UART chips such as 16550, 16650, 16750 or 16950.
  
-- Use the RS232 **cable** that was supplied with the system. When required to obtain a new one, check it is a 'straight forwarded' type. Do not use null-modem cables.+- Use the RS232 **cable** which was supplied with the system. When required to obtain a new one, check that it is a 'straight forwarded' type. Do not use null-modem cables.
  
 - The software runs under **Windows** but has not been completely tested for Windows 8. Under Windows 7 needs to the compatibility to be set to 'Win XP SP3'. - The software runs under **Windows** but has not been completely tested for Windows 8. Under Windows 7 needs to the compatibility to be set to 'Win XP SP3'.
d230info.txt · Last modified: 23/02/2015 15:28 by pieter