 =====Frequently Asked Questions (Ion-selective electrodes)===== =====Frequently Asked Questions (Ion-selective electrodes)=====
 ====How often do you need to calibrate?==== ====How often do you need to calibrate?====
-Recalibrate at least once a week, or daily if in doubt of your results being accurate or reproducible. Check the manual for more information on each ISE about the expected stability.+Calibrate at least once a week, preferred is **daily**. Calibrate if in doubt of your results being accurate or reproducible. When subject to temperature changes, mind that a difference of 1°C causes a measurement error of 2%. Check the manual for more information on each ISE about the expected stability. 
 ====How long after being opened or made are standards good?==== ====How long after being opened or made are standards good?====
 The stock standard will last at least six months before discarding, whereas diluted standards treated with ISA/pH buffer should be prepared weekly. The stock standard will last at least six months before discarding, whereas diluted standards treated with ISA/pH buffer should be prepared weekly.