=====Frequently Asked Questions (Ion-selective electrodes)===== | =====Frequently Asked Questions (Ion-selective electrodes)===== | ||
- | ===How often do you need to calibrate? | + | ====How often do you need to calibrate?==== |
- | Recalibrate | + | Calibrate |
- | ===How long after being opened or made are standards good?=== | + | ====How long after being opened or made are standards good?==== |
The stock standard will last at least six months before discarding, whereas diluted standards treated with ISA/pH buffer should be prepared weekly. | The stock standard will last at least six months before discarding, whereas diluted standards treated with ISA/pH buffer should be prepared weekly. | ||
- | ===How do you store the electrodes? | + | ====How do you store the electrodes?==== |
It is best to store them dry when they are not to be used in the next week or so. Empty out the filling solution in the gas-sensing electrode. For shorter periods, store in dilute standard approximating the sample concentration, | It is best to store them dry when they are not to be used in the next week or so. Empty out the filling solution in the gas-sensing electrode. For shorter periods, store in dilute standard approximating the sample concentration, | ||
- | ===Can you do temperature compensation with ISE?=== | + | ====Can you do temperature compensation with ISE?==== |
Yes, it is possible, but somewhat difficult. | Yes, it is possible, but somewhat difficult. | ||
- You have to know the isopotential point for a given electrode system. | - You have to know the isopotential point for a given electrode system. | ||
Line 13: | Line 13: | ||
- Very little isopotential point data for ISE's is available at this time. | - Very little isopotential point data for ISE's is available at this time. | ||
**It is best to standardize and measure samples at the same temperature without using temperature compensation.** | **It is best to standardize and measure samples at the same temperature without using temperature compensation.** | ||
- | ===Can you do in-line continuous ISE measurement without treating the sample?=== | + | ====Can you do in-line continuous ISE measurement without treating the sample?==== |
Yes, direct measurement is possible in many cases without ISA/buffer addition to the sample stream. However, Fluoride, Sulfide, Ammonia, and Sodium Electrodes do require pH adjustment and must have ISA/ buffer added to the sample stream. | Yes, direct measurement is possible in many cases without ISA/buffer addition to the sample stream. However, Fluoride, Sulfide, Ammonia, and Sodium Electrodes do require pH adjustment and must have ISA/ buffer added to the sample stream. | ||
- | ===Which standards should be used with the ISE?=== | + | ====Which standards should be used with the ISE?==== |
The most obvious choice will be determined by what | The most obvious choice will be determined by what | ||
concentration units are desired (e.g. ppm as what?). Also, if | concentration units are desired (e.g. ppm as what?). Also, if | ||
Line 21: | Line 21: | ||
sulfate with a lead electrode), use a stock standard of the | sulfate with a lead electrode), use a stock standard of the | ||
ion to be measured (e.g. sulfate). | ion to be measured (e.g. sulfate). | ||
- | ===Why buy a combination electrode instead of separate ones?=== | + | ====Why buy a combination electrode instead of separate ones?==== |
Advantages: | Advantages: | ||
* no external reference electrode needed | * no external reference electrode needed |