Line 12: | Line 12: | ||
- The temperature of the sample can not exceed the operational temperature range of the ISE. | - The temperature of the sample can not exceed the operational temperature range of the ISE. | ||
- Very little isopotential point data for ISE's is available at this time. | - Very little isopotential point data for ISE's is available at this time. | ||
- | It is best to standardize and measure samples at the same temperature without using temperature compensation. | + | **It is best to standardize and measure samples at the same temperature without using temperature compensation.** |
===Can you do in-line continuous ISE measurement without treating the sample?=== | ===Can you do in-line continuous ISE measurement without treating the sample?=== | ||
Yes, direct measurement is possible in many cases without ISA/buffer addition to the sample stream. However, Fluoride, Sulfide, Ammonia, and Sodium Electrodes do require pH adjustment and must have ISA/ buffer added to the sample stream. | Yes, direct measurement is possible in many cases without ISA/buffer addition to the sample stream. However, Fluoride, Sulfide, Ammonia, and Sodium Electrodes do require pH adjustment and must have ISA/ buffer added to the sample stream. | ||
Line 21: | Line 21: | ||
sulfate with a lead electrode), use a stock standard of the | sulfate with a lead electrode), use a stock standard of the | ||
ion to be measured (e.g. sulfate). | ion to be measured (e.g. sulfate). | ||
- | ===Why buy a combination electrode instead of seperate | + | ===Why buy a combination electrode instead of separate |
- | Advantages: no external reference electrode needed more economical than price of both one less electrode to fit in the process. | + | Advantages: |
+ | * no external reference electrode needed | ||
+ | * more economical than price of both | ||
+ | * one less electrode to fit in the process. | ||