To Consort website  |  Print page

User Tools

Site Tools


c10xx_manual

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Last revision Both sides next revision
c10xx_manual [11/06/2014 06:28]
Jan De Ceuster created
c10xx_manual [16/06/2014 08:10]
Jan De Ceuster [Inputs]
Line 19: Line 19:
 The measuring electrodes should be connected to the corresponding coaxial pH/mV or EC connectors. Automatic temperature compensation and temperature measurements are possible by plugging a Pt1000 temperature probe into the °C terminals. Without Pt1000, the manual temperature compensation is automatically switched on. The measuring electrodes should be connected to the corresponding coaxial pH/mV or EC connectors. Automatic temperature compensation and temperature measurements are possible by plugging a Pt1000 temperature probe into the °C terminals. Without Pt1000, the manual temperature compensation is automatically switched on.
  
-Never immerse a pH electrode and a dissolved oxygen electrode together in the same solution! +  * Never immerse a pH electrode and a dissolved oxygen electrode together in the same solution! 
- +  A blinking decimal point warns you for unstable measurements. Wait to read the display! Stirring the solution during the measurements promotes the homogeneity and is obligatory! 
-A blinking decimal point warns you for unstable measurements. Wait to read the display! Stirring the solution during the measurements promotes the homogeneity and is obligatory! +  The instrument will refuse automatic calibration when the electrode is unstable. Insufficient stirring or a worn electrode may be the cause.
- +
-The instrument will refuse automatic calibration when the electrode is unstable. Insufficient stirring or a worn electrode may be the cause.+
 ---- ----
 =====Device Menu===== =====Device Menu=====
c10xx_manual.txt · Last modified: 16/06/2014 08:10 by Jan De Ceuster